| Application | 발전소, 반도체, 케미칼, 석유 화학, 제지 |
| Parameter | Dissolved Oxygen |
| Measurement Range | 0.01 μg/L to 9,999 μg/L |
| Limit of Detection | 0.1 ppb |
| Calibration | Double Known Addition, Offline calibration |
| Temperature Range | -10...120 ˚C |
| Accuracy | ±5 % or 0.1 ppb |
| Pressure | 0.55...6.89 bar (8...100 psig) |

반도체

발전소

석유화학

케미칼

제지
Features
1816DO Features
|
Specification
Applications | • Boiler Feed Water • Ultra Pure Water • Process Steam • Deaerator Outlet • Condensate • Oxygen Feed Control for Plants Using Oxygenated Treatment (OT) |
Parameter | Dissolved Oxygen |
Measurement Range | 0.01 μg/L to 9,999 μg/L |
Accuracy | Dissolved Oxygen: ±2% Reading or 0.1 μg/L Temperature: ±0.1 °C |
| Precision | Dissolved Oxygen: ±2% Reading or 2 Digits Temperature: ±0.1 °C |
Response Time | 90% within 30 sec (Default), Function of Flow and Temperature |
Temperature Range | 2...45 ˚C |
Temperature Compensation | -5.0...105 °C |
Sample Conditions | Flow: 50 mL/min to 200 mL/min Temperature: 2 °C to 45 °C Pressure: < 4 bar) / Drain: Atmospheric |
Output | 2 x 4...20 mA or 0...20 mA Max. load 600 Ω; Convertible: VDC to VDC or 0 VDC to 5 VDC 0...20 mA or 4 ...20 mA |
Display | Custom Backlit LCD with Temperature, Concentration, mV, Error Codes and Menu Driven Prompts |
Dimension | 460 x 300 x 230 mm |
Electrical Supply | 115/230 VAC, 0.25A, 50/60 Hz |